|
These graphs illustrate capacitance in picofarads per foot (pF/ft.),
in various vessel diameters, generated by standard probes offered by
Robertshaw. In the graphs, the sensing probe is calculated as being mounted in
the center of a cylindrical vessel of the diameter shown. The diameters shown
in these graphs are: 4in., 14 in., and 60 in. The probe capacitance values can
be approximated for other vessel diameters by interpolation from the values
obtained from these graphs/diameters.
To calculate picofarad values for any product level, the
following values must be known:
-
Vessel diameter
-
Product dielectric value
-
Probe type and length
-
Desired ZERO & 100% points or the ZERO & DIFFERENTIAL points.
The procedure for utilizing the dielectric constant graphs for
making calculations for a given installation is as follows:
Example: Assumed data
-
Vessel diameter is 14 in.
-
Product dielectric is 60.
-
Probe type is Model 740A-B072 (72 in. active length).
-
Desired ZERO point is 12 in. above the bottom of the active probe end and the
100% level point is 12 in. below the gland. Therefore, a 4 ft. level change
occurs from ZERO to 100% (FULL).
CALCULATIONS
-
At a dielectric of 1 (air), the 740A-B graph shows the pF/ft. is approximately
4.5.
-
At a dielectric of 60, the graph shows the pF/ft. to be approx. 53.
-
With product level up to the ZERO level point (remainder of the probe is in the
air).
-
Immersed section = (1 ft. x 53)=53pF
-
Remainder of probe (5 ft. x 4.5) = 22.5 pF
-
Gland value = 24 pF
-
Total ZERO value = 99.5 pF
-
With the product level at the 100% (FULL) point (Lower 5’ in product and the
upper 1’ in air).
-
Immersed section value (5 ft. x 53) = 265 pF
-
Remainder of probe (1 ft. x 4.5) = 4.5 pF
-
Gland value = 24 pF
-
Total 100% value = 293.5 pF
-
The total CAPACITANCE CHANGE is 293.5 minus 99.5 = 194 pF. For quick
calculation of CAPACITANCE CHANGE value only, subtract AIR pF/ft. (example 4.5)
from PRODUCT pF/ft. (example 53) and multiply by the level difference in feet.
Example: 53 minus 4.5 = 48.5. 48.5 x 4 = 194 pF.
For calculation of probes near an “infinite” flat wall or for probes
mounted in a large cylindrical vessel where the probe is near the wall of the
vessel, calculate the value from the appropriate probe graph using the graph
vessel diameter as twice the distance the probe is from the flat wall or vessel
wall. Take this value and multiply by .77 (77%). This is not exact, but is
close enough for all probe picofarad calculations.
GENERAL NOTES:
-
Conductive Coating Applications:
CAUTION - conductive coatings on the probe will cause an error
in measurement. ALWAYS use anti-coating transmitters for continuous measurement
and shortstop probes/instruments for point level measurement where products
will leave a conductive coating (see “special features” in selection guide
matrix). For continuous measurement with anti-coating transmitters, always use
a probe or probe mounting that provides the highest pF/ft. gain but does NOT
EXCEED the span of the instrument. Gain can be increased by using our high gain
741A probe by mounting the probe close to the wall of the tank or vessel.
-
Dielectric/conductivity changes in product:
For continuous measurement of products that change in
dielectric/conductivity, DO NOT USE any of our capacitance type instruments.
For point level only (no adjustable differential), an ON/OFF instrument with a
horizontally mounted probe or Proximity Plate can be used, provided the
instrument is calibrated using the lowest dielectric/conductivity the process
will ever have.
-
Where a tank or vessel is non-metallic or lined, a metal process ground must be
provided. This can be accomplished by installing a ground rod, using a
concentric shield or ground wire wrapped probe (free flowing, non-viscous
products only), or installing another bare probe to serve as the ground rod.
The least preferred choice is the ground wire wrap. The ground rod should be
installed in close proximity and parallel to the probe in low
dielectric/non-conductive products. For high dielectric/conductive products, it
can be installed anywhere, as long as it is not above the tip of the measuring
probe.
-
When installing a probe into a vessel through a pipe or nozzle, always sheath
the portion of the probe going through the pipe/nozzle (unless a shortstop
probe is used). If a shortstop probe is used (ON/OFF point level only),
the shortstop length (not overall length) must extend a minimum of 4 to 6 in.
(preferably 6 in.) out of the pipe/nozzle and into the tank or vessel (see P.S.
Sheet for 727A/732A for required length.)
NOTE: Only Models 318A and 5100 with shortstop option can be
used with 727A/732A probes.
-
Concentric Shield Option of Probes:
This option can be used for increasing the probe gain in low
dielectric materials, linearizing the LEVEL (not volume) signal in a horizontal
or non-symmetrical tank, or providing a process ground in a non-metal tank.
CAUTION: concentric shields can only be used in free flowing,
non-viscous liquids. Concentric shields are always the same length as the
probe.
-
Flumes, Weirs, and non-cylindrical or horizontal tanks:
Model 5000A with the “Smart Chip” option or Model 7000 or 7000A can
be used for measurement of flow in flumes and weirs, or volume in
non-cylindrical or horizontal tanks. Because the Model 5000A is programmed at
the factory, flume/weir size/type/minimum-maximum flow must be submitted at
order time. For tanks, a strapping table is preferred; otherwise, submit
complete dimensions of tank including type ends/dimensions. The Model 7000 and
7000A are field programmable.
|